Setting Inlet Thresholds

You can set the inlet thresholds so that the alerts are generated when the inlet current and/or voltage crosses the thresholds.

  1. Click Inlet I1 in the PX Explorer pane, and the Inlet I1 page opens in the right pane.
  2. Click Setup. The Inlet I1 Setup dialog appears.
  3. In the Threshold Configuration table, click the sensor whose thresholds you want to configure.
  4. Click Edit or double-click the desired sensor. A threshold setup dialog appears.
  5. Configure the Lower Critical, Lower Warning, Upper Warning and Upper Critical thresholds respectively.
  6. To set the deassertion hysteresis, type a numeric value in the Deassertion Hysteresis field. See "To De-assert" and Deassertion Hysteresis.
  7. To set the assertion timeout, type a numeric value in the Assertion Timeout (samples) field. See "To Assert" and Assertion Timeout.
  8. Click OK in the threshold setup dialog to retain the changes.
  9. To set the thresholds for other sensors, repeat Steps 4 to 8.
  10. Click OK.

Important: The final step is required or the threshold changes are not saved.

See Also

Setting Power Thresholds